• Skip to main content
  • Skip to header right navigation
  • Skip to site footer

786-300-3183 | [email protected]

  • Twitter
  • Facebook
Florida Society of Anesthesiologists

Florida Society of Anesthesiologists

  • About FSA
    • FSA Leadership
      • FSA Past Presidents
      • Distinguished Service Award Past Recipients
      • Recipients of the FSA Presidential Engagement Award
    • FSA Staff
    • FSA NEWS
    • Calendar of Events
    • Contact FSA
    • FSA Charter & Bylaws
    • FSA Speakers Bureau
  • FSA Annual Meeting
    • 2025 Annual Meeting
    • 2024 Annual Meeting Recap
    • Call For Abstracts
    • Past Posters
      • 2025 FSA Podium and Poster Abstracts
      • 2024 FSA Podium and Poster Abstracts
      • 2023 FSA Podium and Poster Abstracts
      • 2022 FSA Podium and Poster Abstracts
      • 2021 FSA Posters
      • 2020 FSA Posters
      • 2019 FSA Posters
      • 2018 FSA Posters
    • Past Meetings
      • 2023 Meeting Recap
      • 2022 Annual Meeting Recap
      • 2019 Annual Meeting Recap
      • 2018 Annual Meeting Recap
  • FSAPAC
    • Donate to the FSAPAC
  • Member Login
  • Member Portal
  • Become a Member
    • FSA Membership Renewal
    • Join the Florida Society of Anesthesiologists (FSA)

2019 FSA Posters

2019 FSA Posters

P002: NEEDLE TIP RECOIL FOLLOWING NEEDLE ADVANCE: A POSSIBLE MECHANISM FOR FALSE LOSS OF RESISTANCE?
Samsun Lampotang, PhD, Brandon Lopez, MD, Andre Bigos, David Lizdas, BSME, Anthony DeStephens, MSME, Terrie Vasilopoulos, PhD, Nikolaus Gravenstein, MD, Michael Cometa, MD; University of Florida

Introduction: Loss of resistance (LOR) is the most common technique to identify the epidural space. However, when using fluoroscopy to confirm the accuracy of needle placement within the epidural space, incidence of false LOR can be high, e.g., 74.7% during cervical interlaminar epidural injections at C7-T1. To distinguish true LOR from false LOR, techniques like applying pressure with two fingers at the needle entry point have been devised (bi-digital pressure test; BiP Test). We made an incidental observation (needle tip recoil) during a simulator-based study that may be a potential mechanism for false LOR.

Methods: After IRB approval and informed consent, we studied three LOR techniques on a custom-built mixed reality simulator with 45 providers: Incremental needle advancement, with Intermittent LOR assessment (II); Continuous needle advancement, with high-frequency Intermittent LOR assessment (CI); and Continuous needle advancement, with Continuous LOR assessment (CC). Participants pushed a needle tracked with a magnetic sensor (sub-millimeter resolution) near its tip into gel simulating human tissue. LOR is simulated by allowing the plunger to slide forward if tapped and can be felt only when the needle tip position is past a virtual LOR plane. Each provider identified LOR using each technique 5 times with LOR set at random depths in the simulator software. To assess differences among the 3 LOR techniques in occurrence of false LOR, Cochran’s Q test was used, which is a repeated measures analysis for dichotomous outcomes. This accounts for potential within-participant correlations among the techniques. We replaced the gel block with a lower spine of a freshly slaughtered pig with a fat layer of ~20 mm to determine if needle recoil is present in animal tissue as well.

Results: We unexpectedly observed needle recoil with all 3 techniques: II, CI, and CC, averaging 2.5 to 3.5 and as high as 5 mm (Fig. 1). Needle recoil when releasing pressure on the syringe can cause the needle tip to retract behind the virtual LOR plane after initially obtaining LOR, thus creating a false LOR position (Fig. 1). There was a higher likelihood of the final overshoot position retaining LOR with the II technique versus CC or CI (χ2=7.9, df=2, P=0.019; Fig 2). We obtained similar magnitudes (average ~3 mm, as high as 5 mm) of needle tip recoil in animal tissue (Fig. 1) if the needle initially traversed the fat layer, as those obtained in gel.  

Figure 1. Needle advancement over time for various advancement techniques in ballistics gel and pig tissue medium.

Figure 2. Percentage of participants who retained LOR at final needle depth. 

Conclusion:  If our simulator and pig needle tip recoil data are representative of what occurs in actual patients, our findings suggest a potential new mechanism for explaining false LOR and new ways to clinically address false LOR. Our data seem to support performing a second LOR assessment before advancing the catheter to verify if the initial LOR is still present, especially with a continuous needle advancement technique. Our data requires verification in a clinical study with actual patients.